SECO 78 Tube Tester




Select or type a tube name to get specific settings for the Eico 635 Tube Tester.


Sort columns values by clicking their titles - A → Z or Z → A

ID Tube IC Fil. Volts R Load L Notes
551 6ES6 6 A 40
552 6ES8 6 B 45
553 6ES8 6 C 45
554 6ET6 6 A 40
555 6EU8 6 B 35
556 6EU8 6 E 35
557 6EW6 6 A 30
558 6EX6 6 D 45 X
559 6EY6 6 D 40 X
560 6F5 6 H 70 X
561 6F6 6 D 80 X
562 6F8 6 H 80 X
563 6F8 6 D 80 X
564 6FH8 6 D 45 X
565 6FM8 6 F 65
566 6FM8 6 B 40
567 6FM8 6 D 40
568 6FV6 6 A 40
569 6FV8 6 A 35
570 6FV8 6 E 35
571 6FY8 6 A 35
572 6FY8 6 G 35
573 6G6 6 D 70 X
574 6G7 6 H 60 X
575 6GH8 6 B 35
OPERATING INSTRUCTIONS
BEFORE YOU TEST
1. Set the "Function Switch" to its mid-position.
2. Set the Switch "L" to position marked "NORMAL"
3. Push the slide switch to "ON" position.
4. Allow a half minute for the tester warm up.
5. The meter pointer should rise to renter scale to the mark labeled "Line Adjust". Adjust control on the panel marked "Line Adjust" with a small screwdriver, if necessary.
TESTING
6. CAUTION: DO NOT INSERT THE TUBE INTO ITS SOCKET UNTIL ALL CONTROLS ARE PROPERLY SET UP.
7. Locate the tube to be tested on the set-up card. Example: Tube Type = 6AU4 -> Fil. V. = 6 -> Sw. "R" = D -> Load = 50 -> Sw. "L" = Leave in "normal" unless indicated by X, Y, or Z
8. Set up the controls as indicated.
9. Insert the tube in its socket. There-are two nine-pin sockets. The one at the tipper left is only used when indicated in the tube set-up.
10. Place the cap clip on top of the tube if one is used.
11. As the tube warms up the meter pointer will gradually read into the "GOOD" area on the right side of the scale. REJECT THE TUBE IF IT READS IN THE RED "REPLACE" area.
12. Push the "Function Switch" forward to the position marked "GCT SHORTS AND LEAKAGE" and tap the tube gently. A good tube should read in the "GOOD" area at the left end of the meter scale. REJECT THE TUBE IF THE POINTER JUMPS INTO THE RED AREA OR STAYS IN THE RED AREA.
13. Tubes having two or more entries on the set-up require more than one test. Set controls as indicated and proceed to test again. NOTE: on many types having two entries it is only necessary to move the SELECTOR SWITCH "R".
14. If it is desired to check a tube for filament continuity ONLY then pull the "FUNCTION" switch to position marked "FILAMENT CONTINUITY". The meter should read in the "GOOD" on the right side of the meter scale if continuity exists.
15. If an asterisk (*) appears opposite the tube set-up, this means that an inter-connection exists in the tube which cannot be isolated with lever switch "L". Such interconnections will indicate the same as a short between these pins and cause the meter to read in the red. Consult the tube manual if necessary.